@article{bcd41e3f90a949e4b49397c4cf02f4d3,
title = "Dosimetric verification of the dynamic intensity-modulated radiation therapy of 92 patients",
author = "Jen-San Tsai and Wazer, \{David E.\} and Ling, \{Marilyn N.\} and Wu, \{Julian K.\} and Marcio Fagundes and Thomas DiPetrillo and Bradley Kramer and Melita Koistinen and Engler, \{Mark J.\}",
note = "Methods and Materials: Anthropomorphic phantoms are used to measure IMRT doses. Four types of verification are developed for: I) system commissioning with beams optimized to irradiate simulated targets in phantoms, II) plans with patient-optimized beams directed to phantoms simulating the patient, III) patient-phantom hybrid plans with patient-optimized beams calculated in phantom without further optimization, and IV) in vivo measurements.",
year = "1998",
month = mar,
day = "1",
doi = "10.1016/S0360-3016(98)00009-1",
language = "American English",
volume = "40",
journal = "International Journal of Radiation Oncology Biology Physics",
}