@article{911af08f58094cbfa986a783fe0609b3,
title = "Determination of machine-specific tolerances using statistical process control analysis of long-term uniform scanning proton machine QA results",
keywords = "monthly QA, proton therapy, statistical analysis",
author = "Suresh Rana",
note = "The 3σ of the dose output, flatness, and symmetry differences in all four rooms were comparable to the TG224 tolerance (±2\%). For the uniform scanning system, if the measured range is compared against the requested range, it may not always be possible to achieve the range difference within ±1 mm (TG ...",
year = "2020",
month = sep,
doi = "10.1002/acm2.12990",
language = "American English",
volume = "21",
journal = "Journal of Applied Clinical Medical Physics",
}